Table 3 RECIST, WHO and volume analyses from US and MR images of Gammex/UTHSCSA Mark 1 phantom

From: Preclinical multimodality phantom design for quality assurance of tumor size measurement

 

Design

Image RECIST and WHO (mean ± SD (% error))

  

US

MRI

RECIST (mm)

17.00

17.29 ± 0.08 (+1.73 ± 0.44%)

16.55 ± 0.64 (-2.65 ± 3.74%)

WHO (mm 2 )

149.00

141.93 ± 1.94 (-4.75 ± 1.30%)

137.73 ± 9.72 (-7.56 ± 6.52%)

 

Design

Image Volume (mean ± SD (% error))

  

US

MRI

Volume (mm 3 )

179.59 (7 mm sphere)

180.45 ± 7.51 (+0.48 ± 4.18%)

152.68 ± 30.81 (-14.99 ± 7.16%)

 

523.59 (10 mm sphere)

498.54 ± 12.60 (-4.78 ± 2.41%)

485.49 ± 31.68 (-7.28 ± 6.05%)

  1. Mean ± standard deviation (SD) for three repeated measures and % errors in parenthesis are shown for RECIST, WHO and volume analysis. RECIST and WHO calculations were based on their definitions. For volume calculation, an equation V=π/6·a·b·c where a, b and c are diameters in three perpendicular dimensions was used.